Dual-level Acoustic Calibrator
Acoustic CalibratorsEnsure your noise measurements are compliant with our dual-level acoustic calibrator. You can calibrate both sound level meters and noise dosimeters at both 94dB and 114dB at 1kHz with one instrument!
- IEC 60942: 2017 Class 1 or Class 2 performance
- Dual-level 94dB and 114dB @ 1kHz sound pressure level outputs
- Suitable for use with a standard 1/2″ microphone capsule
- Ideal for sound level meters and noise dosimeters
The Dual-level Acoustic Calibrator
Class 1 and Class 2 Dual-level Acoustic Calibrators
Our dual-level acoustic calibrator (sound calibrator) is suitable for the field and laboratory calibration of sound level meters, noise dosimeters, and other noise measurement instruments.
The calibrator is simple to operate and meets the requirements of the latest IEC 60942:2017 standard for acoustic calibrators with performance to Class 1 or Class 2 and features both 94dB and 114dB at 1kHz sound pressure level output in one single instrument.
It’s ideal for use with a variety of noise dosimeters, where 94dB or 114dB can be used in preference.
Learn more about our Dual-Level Acoustic Calibrator
If you would like to learn more, a member of our team will be happy to help
The Dual‑Level Acoustic Calibrator is a handheld device used to check and calibrate sound level meters, noise dosimeters and other acoustic measurement equipment by generating consistent reference sound levels. It includes two output levels (94 dB and 114 dB) in a single instrument for more flexible calibration options.
This calibrator produces both 94 dB and 114 dB at 1 kHz, allowing you to verify instrument response at two different reference sound levels – useful for instruments that require checks at higher levels as well as standard field calibration.
It’s suitable for calibrating:
-
Sound level meters
-
Noise dosimeters
-
Other acoustic measurement devices equipped with a standard ½″ microphone capsule.
The Dual‑Level Acoustic Calibrator complies with the IEC 60942:2017 standard and is available with Class 1 or Class 2 performance, depending on the model you choose.
Yes, it’s designed to be simple and intuitive. You just need to attach the calibrator to the microphone under test, press the button to generate the calibration tone, and use the LED indicator to confirm operation.
It runs from a single 9 V battery (PP3/6LR61/6F22), making it easy to use in the field or in the lab without requiring mains power.
The calibrator is designed for standard ½″ microphones per IEC 1094‑4. An adaptor is available for ¼″ microphones if needed.
Both output levels (94 dB and 114 dB) are specified with tight tolerances (typically ±0.2 dB), ensuring reliable reference levels for instrument verification.
Regular calibration ensures your noise measurement equipment remains accurate and compliant with standards, which is essential for providing trustworthy and defensible data in workplace assessments, environmental studies, or compliance reporting.
Applicable standards
CR:516 IEC 60942:2017 Class 2
CR:517 IEC 60942:2017 Class 1
Sound pressure level (re 20uPa)
94dB +/- 0.2dB
114dB +/- 0.2dB
Frequency
1kHz +/- 1%
Distortion
<2%
Reference conditions
23oC, 101.3kPa, 65% RH
Microphone size
1/2″ (to IEC 1094-4)
Calibration adaptor
QD:7 for use with 1/4″ microphone capsules
Battery type
1 x 9v (PP3, 6LR61, 6F22)
Dimensions
121mm x 47mm diameter
Weight
220 gms with battery
Controls
Push button start, Automatic switch off with green LED indication. User override of automatic switch off
CE classification
EN61000-6-3:2007+A1:2011
EN61000-6-1:2007




There are no reviews yet.