We’re delighted to announce that the optimus red and optimus green sound level meters have been granted Type Approval to UNE-EN 61672 Parts 1 & 2 in Spain by Applus+, one of the world’s leading testing, inspection and certification companies.
After a series of long and detailed tests, which include both electrical and acoustic verification, approval numbers have been issued for the Class 1 optimus green and optimus red instruments as follows:
CR:161A, CR:161B & CR:161C
CR:1710, CR:171A & CR:171B
The approval numbers are:
No. 02-001-B-62/13-R the CR:161 (optimus red) sound level meters
No. 02-001-B-63/13-R CR:171 (optimus green) sound level meters
The standards that were used in the verification were:
UNE-EN 61672-1: electroacºstica : son³metros. Parte 1, Especificaciones
UNE-EN 61672-2: electroacºstica : son³metros. Parte 2, Ensayos de evaluaci³n de modelo
What is Type Approval?
Type Approval, or Pattern Approval, is a method where a test laboratory such as the Applus, PTB or LNE tests an instrument against a set of standard tests to ensure that it meets the performance claims of the manufacturer.
The optimus sound level meters have been designed to meet the requirements of IEC 61672 (and the corresponding UNE-EN, EN and DIN standards) and so these independent test have been carried out to verify that that the instruments do indeed meet these requirements. The IEC 61672 standard is in three parts.
The first of these is known as Electroacoustics Sound level meters Part 1: Specifications and this defines the performance specifications for the instruments.
Part 2 is known as Electroacoustics Sound level meters Part 2: Pattern evaluation tests and this defines the tests and checks that a laboratory must carry out when they are testing an instrument for Type or Pattern approval. These are the tests that the PTB have been carrying out on the optimus sound level meters.
Part 3 is known as Electroacoustics Sound level meters Part 3: Periodic tests and this defines the tests that should be carried out during a Periodic or routine verification.
This post has more information about the different sections of the IEC 61672 standard.
Related Posts:
What is Type or Pattern Approval?
PTB Type Approval for the Optimus Sound Level Meters
Tonal Noise Detection with the optimus green sound level meters