LED Indicators on the doseBadge5

The doseBadge5 uses a set of LED indicators to show the status of measurements, charging, calibration and general operation.

Four LEDs (orange, green, blue and red) are used in combination to show the current status and activity of the instrument as follows.

Where an LED is flashing, this is shown as Flash Time/Period. For example, 0.1/2.0 is one short flash of 0.1 second every 2 seconds and 0.1/5.0 is one short flash of 0.1 second every 5 seconds.

doseBadge5 on the Charging Dock

StatusLeft LEDRight LEDNotes
Badge chargingSolid orange
Badge fully chargedSolid green
Badge charging and measuringSolid orangeFlashing blue (0.1/2.0)
Badge charged and measuringSolid greenFlashing blue (0.1/2.0)
Badge charging and memory full (or badge error)Solid orangeSolid red
Badge charged and memory full (or badge error)Solid greenSolid red
Calibration successfulSolid blue (5 seconds)The badge will revert to the previous indication after 5 seconds
Calibration failedSolid redThe badge will revert to the previous indication after 5 seconds
Instrument start-upSolid orangeApproximately 1 second

doseBadge5 undocked

Status Left LED Right LED Notes
Badge on (not measuring) Flashing green (0.1/2.0)    
Badge measuring   Flashing blue (0.1/2.0)  
Badge measuring & dose threshold exceeded   Flashing red (0.1/2.0)  
Badge paused   Flashing blue (0.1/5.0)  
Badge paused & dose threshold
exceeded
  Flashing red (0.1/5.0)  
Battery low (not measuring) Flashing orange (0.1/2.0)    
Battery low & measuring Flashing orange (0.1/2.0) Flashing blue (0.1/2.0) Alternating flashes
Battery low & paused Flashing orange (0.1/5.0) Flashing blue (0.1/5.0) Alternating flashes
Battery low, measuring & dose
threshold exceeded
Flashing orange (0.1/2.0) Flashing red (0.1/2.0) Alternating flashes
Battery low, paused & dose
threshold exceeded
Flashing orange (0.1/5.0) Flashing red (0.1/5.0) Alternating flashes
Calibration successful   Solid blue (5 seconds) The badge will revert to the
previous indication after 5 seconds
Calibration failed   Solid red (5 seconds) The badge will revert to the
previous indication after 5 seconds
Instrument start-up   Solid orangeApproximately 1 second

Error conditions or warning indications

StatusLeft LEDRight LEDNotes
Memory fullSolid redConnect to NoiseTools to download & clear measurements
Badge errorSolid redConnect to NoiseTools to confirm status